Fluke 753 Overview
You’re going to love what we’ve done with your calibrator.
The 753 is everything you needed and wished for in the 743B and more!
The 753 does the work of several tools – sourcing, simulating and measuring pressure, temperature, and electrical signals in one rugged, hand-held device. You’ll fall in love with the new improved graphical screen, Li-Ion battery for longer life, USB port and new accessories to complete the package.
For documentation, the 753 automates calibration procedures and captures your data. And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations.
Want HART communication built into your 753 calibrator? Consider the Fluke 754 Documenting Process Calibrator with HART.
Fluke 753 Highlights
- Save over $485 with Fluke 753 VIP Kit.
- Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments
- Source/simulate volts, mA, thermocouples , RTDs, frequency, ohms, and pressure to calibrate transmitters
- Power transmitters during test using loop supply with simultaneous mA measurement
- Measure/source pressure using any of 29 Fluke pressure modules
- Advanced features like autostep, custom units, user entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay and more
- Includes the new C799 Field soft case
- Three-year warranty
>
Each Fluke 753 Includes
- Three sets of stackable test leads
- Three sets of TP220 test probes with three sets of “extended tooth” alligator clips
- Two sets AC280 hook clips
- BP7240 Li-ion battery pack
- BC7240 battery charger
- C799 field soft case
- USB communication cable
- Getting started guide
- Instruction manual on CDROM
- NIST traceable certificate of calibration
- Rugged Case
>
Measurement Accuracy | |||||
Voltage DC | Range /Resolution | 1 Year | 2 Years | ||
100.000 mV | 0.02% 0.005 mV | 0.03% 0.005 mV | |||
3.00000 V | 0.02% 0.00005 V | 0.03% 0.00005 V | |||
30.0000 V | 0.02% 0.0005 V | 0.03% 0.0005 V | |||
300.00 V | 0.05% 0.05 V | 0.07% 0.05 V | |||
Voltage AC | 3.000 V (40 Hz to 500 Hz) / 0.001 V | 0.5% 0.002 V | 1.0% 0.004 V | ||
30.00 V (40 Hz to 500 Hz) / 0.01 V | 0.5% 0.02 V | 1.0% 0.04 V | |||
300.0 V (40 Hz to 500 Hz) / 0.1 V | 0.5% 0.2 V | 1.0% 0.2 V | |||
Current DC | 30.000 mA | 0.01% 5 uA | 0.015% 7 uA | ||
110.00 mA | 0.01% 20 uA | 0.015% 30 uA | |||
Resistance | 10.000 Ω | 0.05% 50 mΩ | 0.07% 70 mΩ | ||
100.00 Ω | 0.05% 50 mΩ | 0.07% 70 mΩ | |||
1.0000 kΩ | 0.05% 500 mΩ | 0.07% 0.5 Ω | |||
10.000 kΩ | 0.1% 10 Ω | 0.15% 15 Ω | |||
Frequency | 1.00 to 110.00 Hz / 0.01 Hz | 0.05 Hz | |||
110.1 to 1100.0 Hz / 0.1 Hz | 0.5 Hz | ||||
1.101 to 11.000 kHz / 0.001 kHz | 0.005 kHz | ||||
11.01 to 50.00 kHz / 0.01 kHz | 0.05 kHz | ||||
Source Accuracy | |||||
1 Year | 2 Years | ||||
Voltage DC | 100.000 mV | 0.01% 0.005 mV | 0.015% 0.005 mV | ||
1.00000 V | 0.01% 0.00005 V | 0.015% 0.0005 V | |||
15.0000 V | 0.01% 0.0005 V | 0.015% 0.0005 V | |||
Current DC | 22.000 mA (source) | 0.01% 0.003 mA | 0.02% 0.003 mA | ||
Current sink (simulate) | 0.02% 0.007 mA | 0.04% 0.007 mA | |||
Resistance | 10.000 Ω | 0.01% 10 mΩ | 0.015% 15 mΩ | ||
100.00 Ω | 0.01% 20 mΩ | 0.015% 30 mΩ | |||
1.0000 kΩ | 0.02% 0.2 Ω | 0.03% 0.3 Ω | |||
10.000 kΩ | 0.02% 3 Ω | 0.03% 5 Ω | |||
Frequency | 0.1 to 10.99 Hz | 0.01 Hz | |||
0.01 to 10.99 Hz | 0.01 Hz | ||||
11.00 to 109.99 Hz | 0.1 Hz | ||||
110.0 to 1099.9 Hz | 0.1 Hz | ||||
1.100 to 21.999 kHz | 0.002 kHz | ||||
22.000 to 50.000 kHz | 0.005 kHz | ||||
Technical Data | |||||
Data log functions | Measure functions | Voltage, current, resistance, frequency, temperature, pressure | |||
Reading rate | 1, 2, 5, 10, 20, 30, or 60 readings/minute | ||||
Maximum record length | 8000 readings (7980 for 30 or 60 readings/minute) | ||||
Ramp functions | Source functions | Voltage, current, resistance, frequency, temperature | |||
Rate | 4 steps/second | ||||
Trip detect | Continuity or voltage (continuity detection not available when sourcing current) | ||||
Loop power function | Voltage | Selectable, 26 V | |||
Accuracy | 10%, 18 V minimum at 22 mA | ||||
Maximum current | 25 mA, short circuit protected | ||||
Maximum input voltage | 50 V DC | ||||
Step functions | Source functions | Voltage, current, resistance, frequency, temperature | |||
Manual step | Selectable step, change with arrow buttons | ||||
Autostep | Fully programmable for function, start delay, stepvalue, time per step, repeat | ||||
Environmental Specifications | |||||
Operating temperature | -10°C to 50°C | ||||
Storage temperature | -20°C to 60°C | ||||
Dust/water resistance | Meets IP52, IEC 529 | ||||
Operating altitude | 3000 m above mean sea level (9842 ft) | ||||
Safety Specifications | |||||
Agency approvals | CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993 | ||||
Mechanical and General Specifications | |||||
Size | 136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in) | ||||
Weight | 1.2 kg (2.7 lb) | ||||
Batteries | Internal Battery Pack Li-ion: 7.2 V, 4400 mAh, 30 Wh | ||||
Battery life | >, 8 hours typical | ||||
Battery replacement | Via snap-shut door without opening calibrator, no tools required | ||||
Side port connections | Pressure module connector | ||||
USB Connector to interface to your PC | |||||
Connection for optional battery charger/eliminator | |||||
Data storage capacity | 1 week of calibration procedures results | ||||
90 day specifications | The standard specification interval for the 750 Series are 1 and 2 years. | ||||
Typical 90 day measurement and source accuracy can be estimated by dividing the one year % of reading or % of output specifications by 2. | |||||
Floor specifications, expressed as % of full scale or counts or ohms remain constant. | |||||
Temperature, Resistance Temperature Detectors | |||||
Degrees or % of reading Type (α) | Range °C | Measure °C1 | |||
1 year | 2 year | ||||
100 Ω Pt (385) | -200 to 100 | 0.07°C | 0.14°C | ||
100 to 800 | 0.02% 0.05°C | 0.04% 0.10°C | |||
200 Ω Pt (385) | -200 to 100 | 0.07°C | 0.14°C | ||
100 to 630 | 0.02% 0.05°C | 0.04% 0.10°C | |||
500 Ω Pt (385) | -200 to 100 | 0.07°C | 0.14°C | ||
100 to 630 | 0.02% 0.05°C | 0.04% 0.10°C | |||
1000 Ω Pt (385) | -200 to 100 | 0.07°C | 0.14°C | ||
100 to 630 | 0.02% 0.05°C | 0.04% 0.10°C | |||
100 Ω Pt (3916) | -200 to 100 | 0.07°C | 0.14°C | ||
100 to 630 | 0.02% 0.05°C | 0.04% 0.10°C | |||
100 Ω Pt (3926) | -200 to 100 | 0.08°C | 0.16°C | ||
100 to 630 | 0.02% 0.06°C | 0.04% 0.12°C | |||
10 Ω Cu (427) | -100 to 260 | 0.2°C | 0.4°C | ||
120 Ω Ni (672) | -80 to 260 | 0.1°C | 0.2°C | ||
Source current | Source °C | Allowable current2 | |||
1 year | 2 year | ||||
1 mA | 0.05°C | 0.10°C | 0.1 mA to 10 mA | ||
0.0125% 0.04°C | 0.025% 0.08°C | ||||
500 μA | 0.06°C | 0.12°C | 0.1 mA to 1 mA | ||
0.017% 0.05°C | 0.034% 0.10°C | ||||
250 μA | 0.06°C | 0.12°C | 0.1 mA to 1 mA | ||
0.017% 0.05°C | 0.034% 0.10°C | ||||
150 μA | 0.06°C | 0.12°C | 0.1 mA to 1 mA | ||
0.017% 0.05°C | 0.034% 0.10°C | ||||
1 mA | 0.05°C | 0.10°C | 0.1 mA to 10 mA | ||
0.0125% 0.04°C | 0.025% 0.08°C | ||||
1 mA | 0.05°C | 0.10°C | 0.1 mA to 10 mA | ||
0.0125% 0.04°C | 0.025% 0.08°C | ||||
3 mA | 0.2°C | 0.4°C | 0.1 mA to 10 mA | ||
1 mA | 0.04°C | 0.08°C | 0.1 mA to 10 mA | ||
1. For two and three-wire RTD measurements, add 0.4°C to the specifications. | |||||
2. Supports pulsed transmitters and PLCs with pulse times as short as 1 ms | |||||
Temperature, Thermocouples | |||||
Type | Source °C | Measure °C | Source °C | ||
1 year | 2 years | 1 year | 2 years | ||
E | -250 to -200 | 1.3 | 2 | 0.6 | 0.9 |
-200 to -100 | 0.5 | 0.8 | 0.3 | 0.4 | |
-100 to 600 | 0.3 | 0.4 | 0.3 | 0.4 | |
600 to 1000 | 0.4 | 0.6 | 0.2 | 0.3 | |
N | -200 to -100 | 1.0 | 1.5 | 0.6 | 0.9 |
-100 to 900 | 0.5 | 0.8 | 0.5 | 0.8 | |
900 to 1300 | 0.6 | 0.9 | 0.3 | 0.4 | |
J | -210 to -100 | 0.6 | 0.9 | 0.3 | 0.4 |
-100 to 800 | 0.3 | 0.4 | 0.2 | 0.3 | |
800 to 1200 | 0.5 | 0.8 | 0.3 | 0.3 | |
K | -200 to -100 | 0.7 | 1 | 0.4 | 0.6 |
-100 to 400 | 0.3 | 0.4 | 0.3 | 0.4 | |
400 to 1200 | 0.5 | 0.8 | 0.3 | 0.4 | |
1200 to 1372 | 0.7 | 1 | 0.3 | 0.4 | |
T | -250 to -200 | 1.7 | 2.5 | 0.9 | 1.4 |
-200 to 0 | 0.6 | 0.9 | 0.4 | 0.6 | |
0 to 400 | 0.3 | 0.4 | 0.3 | 0.4 | |
B | 600 to 800 | 1.3 | 2 | 1 | 1.5 |
800 to 1000 | 1 | 1.5 | 0.8 | 1.2 | |
1000 to 1820 | 0.9 | 1.3 | 0.8 | 1.2 | |
R | -20 to 0 | 2.3 | 2.8 | 1.2 | 1.8 |
0 to 100 | 1.5 | 2.2 | 1.1 | 1.7 | |
100 to 1767 | 1 | 1.5 | 0.9 | 1.4 | |
S | -20 to 0 | 2.3 | 2.8 | 1.2 | 1.8 |
0 to 200 | 1.5 | 2.1 | 1.1 | 1.7 | |
200 to 1400 | 0.9 | 1.4 | 0.9 | 1.4 | |
1400 to 1767 | 1.1 | 1.7 | 1 | 1.5 | |
C | 0 to 800 | 0.6 | 0.9 | 0.6 | 0.9 |
800 to 1200 | 0.8 | 1.2 | 0.7 | 1 | |
1200 to 1800 | 1.1 | 1.6 | 0.9 | 1.4 | |
1800 to 2316 | 2 | 3 | 1.3 | 2 | |
L | -200 to -100 | 0.6 | 0.9 | 0.3 | 0.4 |
-100 to 800 | 0.3 | 0.4 | 0.2 | 0.3 | |
800 to 900 | 0.5 | 0.8 | 0.2 | 0.3 | |
U | -200 to 0 | 0.6 | 0.9 | 0.4 | 0.6 |
0 to 600 | 0.3 | 0.4 | 0.3 | 0.4 | |
BP | 0 to 1000 | 1 | 1.5 | 0.4 | 0.6 |
1000 to 2000 | 1.6 | 2.4 | 0.6 | 0.9 | |
2000 to 2500 | 2 | 3 | 0.8 | 1.2 | |
XK | -200 to 300 | 0.2 | 0.3 | 0.2 | 0.5 |
300 to 800 | 0.4 | 0.6 | 0.3 | 0.6 |